Tailor-made infrastructure for real-time investigation of prefilled syringes under varying temperature and pressure conditions using Synchrotron CT
• Capability to measure the pre-filled syringe using Synchrotron CT
• Temperature control of the pre-filled syringe via a liquid cooler
• Rotary union for the cooling liquid and pressure
• Temperature and pressure sensors for measurement and control
• High control accuracy
• High heat transfer from chamber to syringe
• Pressurization via vacuum pump
• Pressure sensor for measuring and controlling the syringe ambient pressure
• Easy sample change
• Sample replacement is possible without deactivating the liquid cooler
• Temperature sensors in the pressure chamber
• Temperature measurement directly on the glass body
• Use of a rotary union with slip rings for cooling fluid (inlet/outlet), pressure, and 2x PT100 temperature sensors
• Rotary union with low torque <= 0.5 Nm
• High heat transfer to syringes via aluminum insert
• Installation and commissioning of hardware and software
• Controller tuning and acceptance testing in the laboratory
• Setup of the in-situ chamber on the beamline
• Conducting Synchrotron CT imaging experiments according to customer-specific requirements for pressure and temperature